User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Honeywell UDCÿ3000:Universal Controller DATE: 02-Jan-96 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 46 ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F 1.002 ASK+ X 1.003 DISP Install UUT in mainframe and apply power. 1.004 DISP Allow the UUT 1 hour warm-up. 1.005 DISP On UUT, select SETUP 5 times for the INPUT 1 display. 1.006 DISP Select FUNC key 3 times for the IN1 TYPE display. 1.007 OPBR Does the UUT display T TC L? 1.008 JMPF 1.011 1.009 CALL Honeywell UDC 3000 Type T øF - Low 1.010 JMP 1.031 1.011 OPBR Does the UUT display T TC H? 1.012 JMPF 1.015 1.013 CALL Honeywell UDC 3000 Type T øF - High 1.014 JMP 1.031 1.015 OPBR Does the UUT display J TC H? 1.016 JMPF 1.019 1.017 CALL Honeywell UDC 3000 Type J øF - High 1.018 JMP 1.031 1.019 OPBR Does the UUT display K TC H? 1.020 JMPF 1.023 1.021 CALL Honeywell UDC 3000 Type K øF - High 1.022 JMP 1.031 1.023 OPBR Does the UUT display 4 - 20mA? 1.024 JMPF 1.031 1.025 DISP Select FUNC key 2 times for the IN1 HI display. 1.026 OPBR Does the UUT display 100.0? 1.027 JMPT 1.030 1.028 CALL Honeywell UDC 3000 HG 1.029 JMP 1.031 1.030 CALL Honeywell UDC 3000 RH 1.031 HEAD {} 1.032 DISP {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 1.032 DISP {º THIS COMPLETES THE VERIFICATION º} 1.032 DISP {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 1.033 END